On December 5, 2017, the Ministry of Science and Technology of Vietnam issued Circular 15/2017/TT-BKHCN regarding the National Technical Regulation for Linear Accelerators used in Radiotherapy.
According to Circular 15, linear accelerators used in radiotherapy must meet the following requirements:
No. | Inspection Content | Technical Requirements |
---|---|---|
I | External Inspection | - The radiotherapy linear accelerator to be inspected must be in working condition. - The radiotherapy linear accelerator and control console must be clean, free of rust, cracks, or deformation. |
II | Technical Inspection | |
1 | Control Console Inspection | - The switches on the control console must function properly and move smoothly; - Indicator lights on the control console must accurately reflect the operating status of the linear accelerator. |
2 | Linear Accelerator Operation Inspection | - The linear accelerator arm must rotate smoothly around its axis and be able to stop at any position. - The beam collimator of the linear accelerator must rotate easily around its axis and stop at any position. - MLC leaves or jaws can move easily under control to create the required radiation field. - The light field size indicator must be clearly visible. - Mechanical distance measurement systems, optical indicators, and laser systems must be clearly visible. - The emergency stop system allows stopping irradiation from the control console, and resuming irradiation after stopping can only be done from the control console. - The interlock system operates stably. |
III | Measurement Inspection | |
A | Mechanical System Accuracy Inspection | |
1 | Accuracy of Optical Indicator Ruler | ± 2 mm |
2 | Accuracy of Rotation Angle Display in the Beam Collimator System, Arm, and Treatment Table | ± 1° |
3 | Congruence of Light Field and Radiation Field Size | ± 3 mm for field sizes ≤ (200 mm x 200 mm); ± 5 mm for field sizes > (200 mm x 200 mm) ± 2 mm deviation of the radiation field center from the crosshair center |
4 | Laser Beam Accuracy at the Isocenter | ± 2 mm |
5 | Accuracy in the Treatment Table Movement | ± 2 mm |
B | Photon Beam Radiation Dose Inspection | |
6 | Photon Beam Quality (TPR20/10) | ± 2% compared to the TPR20/10 or PDD20/10 value at the time of commissioning |
7 | Photon Beam Output Dose Calibration | 1 cGy/MU ± 2% |
8 | Output Dose Factor according to Field Size | ± 2% compared to the output dose factor value at the time of commissioning |
9 | Field Flatness (F) and Symmetry (S) | ± 3% |
10 | Transmission Factor of Wedge (Physical Wedge) | ± 2% compared to the transmission factor value at the time of commissioning |
C | Electron Beam Radiation Dose Inspection | |
11 | Electron Beam Quality (R50) | ± 2 mm compared to the R50 value at the time of commissioning |
12 | Electron Beam Output Dose Calibration | 1 cGy/MU ± 2% |
13 | Output Dose Factor for Electron Beam Applicators | ± 2% compared to the output dose factor value at the time of commissioning |
14 | Field Flatness (F) and Symmetry (S) for Electron Beam | F ≤ ± 6% for energies less than 10 MeV; F ≤ ± 4% for energies greater than 10 MeV S ≤ ± 3% |
D | Multi-Leaf Collimator (MLC) Inspection | |
15 | Accuracy of MLC Leaf Positioning at Isocenter | ± 1 mm compared to the MLC coordinate value established by software |
16 | Congruence of Radiation Fields Created by MLC When Rotating the Gantry (Gantry Spoke Shot) | All intersection points lie within a circle with a radius of 1 mm |
17 | Congruence of MLC-Created Radiation and Light Field Sizes | ± 2 mm |
18 | Congruence of Radiation Fields When Rotating the MLC (MLC Spoke Shot) | All intersection points lie within a circle with a radius of 1 mm |
19 | Radiation Leakage through Leaf Gaps and Body (Interleaf/Intraleaf Leakage), Leakage through Leaf Tips | ≤ 2% (through leaf bodies), ≤ 2.5% (through leaf gaps), and all values of radiation leakage through leaf bodies, gaps, and tips must be incorporated into the radiotherapy planning simulation software |
More details can be found in Circular 15/2017/TT-BKHCN of the Ministry of Science and Technology of Vietnam, effective from December 15, 2018.